01785 600 691enquiries@thecalcentre.co.uk

Fluke 753 Documenting Process Calibrator

£6,826.00 Inc VAT: £8,191.20

    Available on back-order

    Calibration Certificate

    Add a 12 Month Calibration Certificate

    Description

    Fluke 753 Documenting Process Calibrator

    Powerful, multifunctional and condensed down into a portable instrument, the Fluke 753 Documenting Process Calibrator can measure and source everything from temperature to pressure. It can be used in practically any calibration scenario and gives you all of the tools you’ll need in one comprehensive instrument.

    The 753 is truly a multifunction instrument; it can measure voltage, mA, RTDs, thermocouples, frequency and ohms, source/simulate voltage, mA, RTDs, frequency, ohms and pressure and includes dedicated memory for storing up to a full week of downloaded procedures and calibration results.

    With the 753 at hand, you can configure automated calibration procedures for linear transmitters, DP flow transmitters and one or two-point limit switches. This means you only have to select your configuration, connect up and start testing with absolutely minimal effort on your part.

    When measuring pressure, you can use Fluke’s expansive range of pressure modules (sold separately) in combination with this instrument to cater for practically any pressure application ranging from gauge to vacuum.

     

    Our customers are also considering the Fluke 754 with HART Communication

    Measurement Accuracy
    Voltage DCRange /Resolution1 Year2 Years
    100.000 mV0.02% 0.005 mV0.03% 0.005 mV
    3.00000 V0.02% 0.00005 V0.03% 0.00005 V
    30.0000 V0.02% 0.0005 V0.03% 0.0005 V
    300.00 V0.05% 0.05 V0.07% 0.05 V
    Voltage AC3.000 V (40 Hz to 500 Hz) / 0.001 V0.5% 0.002 V1.0% 0.004 V
    30.00 V (40 Hz to 500 Hz) / 0.01 V0.5% 0.02 V1.0% 0.04 V
    300.0 V (40 Hz to 500 Hz) / 0.1 V0.5% 0.2 V1.0% 0.2 V
    Current DC30.000 mA0.01% 5 uA0.015% 7 uA
    110.00 mA0.01% 20 uA0.015% 30 uA
    Resistance10.000 Ω0.05% 50 mΩ0.07% 70 mΩ
    100.00 Ω0.05% 50 mΩ0.07% 70 mΩ
    1.0000 kΩ0.05% 500 mΩ0.07% 0.5 Ω
    10.000 kΩ0.1% 10 Ω0.15% 15 Ω
    Frequency1.00 to 110.00 Hz / 0.01 Hz0.05 Hz
    110.1 to 1100.0 Hz / 0.1 Hz0.5 Hz
    1.101 to 11.000 kHz / 0.001 kHz0.005 kHz
    11.01 to 50.00 kHz / 0.01 kHz0.05 kHz
    Source Accuracy
    1 Year2 Years
    Voltage DC100.000 mV0.01% 0.005 mV0.015% 0.005 mV
    1.00000 V0.01% 0.00005 V0.015% 0.0005 V
    15.0000 V0.01% 0.0005 V0.015% 0.0005 V
    Current DC22.000 mA (source)0.01% 0.003 mA0.02% 0.003 mA
    Current sink (simulate)0.02% 0.007 mA0.04% 0.007 mA
    Resistance10.000 Ω0.01% 10 mΩ0.015% 15 mΩ
    100.00 Ω0.01% 20 mΩ0.015% 30 mΩ
    1.0000 kΩ0.02% 0.2 Ω0.03% 0.3 Ω
    10.000 kΩ0.02% 3 Ω0.03% 5 Ω
    Frequency0.1 to 10.99 Hz0.01 Hz
    0.01 to 10.99 Hz0.01 Hz
    11.00 to 109.99 Hz0.1 Hz
    110.0 to 1099.9 Hz0.1 Hz
    1.100 to 21.999 kHz0.002 kHz
    22.000 to 50.000 kHz0.005 kHz
    Technical Data
    Data log functionsMeasure functionsVoltage, current, resistance, frequency, temperature, pressure
    Reading rate1, 2, 5, 10, 20, 30, or 60 readings/minute
    Maximum record length8000 readings (7980 for 30 or 60 readings/minute)
    Ramp functionsSource functionsVoltage, current, resistance, frequency, temperature
    Rate4 steps/second
    Trip detectContinuity or voltage (continuity detection not available when sourcing current)
    Loop power functionVoltageSelectable, 26 V
    Accuracy10%, 18 V minimum at 22 mA
    Maximum current25 mA, short circuit protected
    Maximum input voltage50 V DC
    Step functionsSource functionsVoltage, current, resistance, frequency, temperature
    Manual stepSelectable step, change with arrow buttons
    AutostepFully programmable for function, start delay, stepvalue, time per step, repeat
    Environmental Specifications
    Operating temperature-10°C to 50°C
    Storage temperature-20°C to 60°C
    Dust/water resistanceMeets IP52, IEC 529
    Operating altitude3000 m above mean sea level (9842 ft)
    Safety Specifications
    Agency approvalsCAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993
    Mechanical and General Specifications
    Size136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in)
    Weight1.2 kg (2.7 lb)
    BatteriesInternal Battery Pack Li-ion: 7.2 V, 4400 mAh, 30 Wh
    Battery life> 8 hours typical
    Battery replacementVia snap-shut door without opening calibrator; no tools required
    Side port connectionsPressure module connector
    USB Connector to interface to your PC
    Connection for optional battery charger/eliminator
    Data storage capacity1 week of calibration procedures results
    90 day specificationsThe standard specification interval for the 750 Series are 1 and 2 years.
    Typical 90 day measurement and source accuracy can be estimated by dividing the one year “% of reading” or “% of output” specifications by 2.
    Floor specifications, expressed as “% of full scale” or “counts” or “ohms” remain constant.
    Temperature, Resistance Temperature Detectors
    Degrees or % of reading – Type (α)Range °CMeasure °C1
    1 year2 year
    100 Ω Pt (385)-200 to 100
    100 to 800
    0.07°C
    0.02% 0.05°C
    0.14°C
    0.04% 0.10°C
    200 Ω Pt (385)-200 to 100
    100 to 630
    0.07°C
    0.02% 0.05°C
    0.14°C
    0.04% 0.10°C
    500 Ω Pt (385)-200 to 100
    100 to 630
    0.07°C
    0.02% 0.05°C
    0.14°C
    0.04% 0.10°C
    1000 Ω Pt (385)-200 to 100
    100 to 630
    0.07°C
    0.02% 0.05°C
    0.14°C
    0.04% 0.10°C
    100 Ω Pt (3916)-200 to 100
    100 to 630
    0.07°C
    0.02% 0.05°C
    0.14°C
    0.04% 0.10°C
    100 Ω Pt (3926)-200 to 100
    100 to 630
    0.08°C
    0.02% 0.06°C
    0.16°C
    0.04% 0.12°C
    10 Ω Cu (427)-100 to 2600.2°C0.4°C
    120 Ω Ni (672)-80 to 2600.1°C0.2°C
    Source currentSource °CAllowable current2
    1 year2 year
    1 mA0.05°C
    0.0125% 0.04°C
    0.10°C
    0.025% 0.08°C
    0.1 mA to 10 mA
    500 μA0.06°C
    0.017% 0.05°C
    0.12°C
    0.034% 0.10°C
    0.1 mA to 1 mA
    250 μA0.06°C
    0.017% 0.05°C
    0.12°C
    0.034% 0.10°C
    0.1 mA to 1 mA
    150 μA0.06°C
    0.017% 0.05°C
    0.12°C
    0.034% 0.10°C
    0.1 mA to 1 mA
    1 mA0.05°C
    0.0125% 0.04°C
    0.10°C
    0.025% 0.08°C
    0.1 mA to 10 mA
    1 mA0.05°C
    0.0125% 0.04°C
    0.10°C
    0.025% 0.08°C
    0.1 mA to 10 mA
    3 mA0.2°C0.4°C0.1 mA to 10 mA
    1 mA0.04°C0.08°C0.1 mA to 10 mA
    1. For two and three-wire RTD measurements, add 0.4°C to the specifications.
    2. Supports pulsed transmitters and PLCs with pulse times as short as 1 ms
    Temperature, Thermocouples
    TypeSource °CMeasure °CSource °C
    1 year2 years1 year2 years
    E-250 to -2001.32.00.60.9
    -200 to -1000.50.80.30.4
    -100 to 6000.30.40.30.4
    600 to 10000.40.60.20.3
    N-200 to -1001.01.50.60.9
    -100 to 9000.50.80.50.8
    900 to 13000.60.90.30.4
    J-210 to -1000.60.90.30.4
    -100 to 8000.30.40.20.3
    800 to 12000.50.80.30.3
    K-200 to -1000.71.00.40.6
    -100 to 4000.30.40.30.4
    400 to 12000.50.80.30.4
    1200 to 13720.71.00.30.4
    T-250 to -2001.72.50.91.4
    -200 to 00.60.90.40.6
    0 to 4000.30.40.30.4
    B600 to 8001.32.01.01.5
    800 to 10001.01.50.81.2
    1000 to 18200.91.30.81.2
    R-20 to 02.32.81.21.8
    0 to 1001.52.21.11.7
    100 to 17671.01.50.91.4
    S-20 to 02.32.81.21.8
    0 to 2001.52.11.11.7
    200 to 14000.91.40.91.4
    1400 to 17671.11.71.01.5
    C0 to 8000.60.90.60.9
    800 to 12000.81.20.71.0
    1200 to 18001.11.60.91.4
    1800 to 23162.03.01.32.0
    L-200 to -1000.60.90.30.4
    -100 to 8000.30.40.20.3
    800 to 9000.50.80.20.3
    U-200 to 00.60.90.40.6
    0 to 6000.30.40.30.4
    BP0 to 10001.01.50.40.6
    1000 to 20001.62.40.60.9
    2000 to 25002.03.00.81.2
    XK-200 to 3000.20.30.20.5
    300 to 8000.40.60.30.6
    • Measure volts, mA, RTDs, thermocouples, frequency and ohms (ideal for testing sensors, transmitters and other instruments)
    • Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms and pressure
    • Power transmitters during test using loop supply with simultaneous mA measurements
    • Compatible with 29 Fluke 700Pxx Pressure Modules (sold separately)
    • Create automated calibration procedures to speed up your testing time
    • Record and document results on built-in memory that can store a full week of download procedures and calibration results
    • Advanced features – autostep, custom units, user-entered values during test, one-point and two-point switch testing, square root DP flow testing and programmable measurement delay
    • Easy to use – minimal configuration options and advanced features make calibration a breeze
    • Rechargeable batteries – 10 hours of use on a full charge
    • Handling of fast-pulsed RTD transmitters and PLCs with pulses as short as 1ms
    • Fluke 753 Documenting Process Calibrator
    • 3 x Sets of Stackable Test Leads
    • 3 x Sets of TP220 Test Probes with 3 x Extended Tooth Alligator Clips
    • 2 x Sets of AC280 Hook Clips
    • BP7240 Li-Ion Battery Pack
    • BC7240 Battery Charger
    • C799 Field Soft Case
    • USB Communication Cable
    • Getting Started Guide
    • Instruction Manual on CD-ROM
    • NIST Traceable Certificate of Calibration
    • DPC/TRACK 2 Sample Software